system, etc.
fluorometry.
Instrument Characteristics
· Measurement mode: automatic measurement mode and
manual measurement mode.
It can choose reasonable measuring gear intelligently
according to the sample content in automatic
measurement mode.
· Wide range. When content is less than 6.0×10-6 g/ml, sample
can be measured directly without dilution operation.
· Display the result of liquid sample or solid sample
according to the the original state of the sample (μg/ml
or μg/g).
· Import ultraviolet source, long lifetime, analyze at least
80000 uranium samples.
· TFT 5 inch color LCD display.
· Equipped with embedded thermal printer, the measured
results can be directly printed.
· Instrument can be connected PC through USB and
support for remote control operation.
Specifications
The instrument meets the requirements of EJ/T 823
standard, the main performance indexes are as follows:
1. Measuring objects: Liquid sample. Samples need to be
prepared and converted to liquid samples
2. Detection Limit: ≤ 0.01 ng/ml
(three times standard deviation)
3. Precision: RSD ≤ 5%
4. Accuracy: MPE ±8%
5. Measuring Range: (0.01×10-9 – 6.0×10-6) g/ml
6. Linearity: Correlation coefficient ≥ 0.9992
7. Stability (8h): RE ≤ 7%
(Uranium concentration: 2.0 ng/ml)
8. Data Print: Thermal micro printer
9. Communication Type: USB
10. Storage Capacity: 4000 sets of measuring data
11. Operating Environment:
Temperature: (+10 – +40)°C
Relative humidity: < 85% (+30°C)
12. Power/Power Consumption: 220V/50Hz, about 50 W
13. Dimension and Weight: (45×36×19) cm 15 kg